Warranty
IJ electrodes have a warranty of 12 months from date of purchase.
Any electrode found to be faulty due to manufacture within
this time will be replaced.
Ionode reserves the right to limit or modify product warranty
if it is deemed that the electrode has been used in unsuitable
applications. Electrodes with broken stems, connectors or
cables are not covered by warranty.
Electrode life may be reduced in chemically aggressive or
abrasive samples and at high temperatures.
Visit ionode.com for more details.
User maintenance and troubleshooting
IJ- F
FLUORIDE
ION SELECTIVE
ELECTRODE
Symptom Possible cause Remedy
Drift Junction blocked Remove and clean sleeve
Membrane unclean
or contaminated
Remove sleeve. Gently
polish membrane with
ne polishing agent
Noisy Poor connection
to meter
Check connection
Junction not
immersed fully
Lower electrode into
solution below junction
Insufcient
electrolyte or
bubble
Rell electrolyte
Inaccurate
readings
or poor
reproduc-
ibility
Contaminated
standard(s)
Make up fresh calibration
standards and recalibrate
Drift since last
calibration
Recalibrate using freshly
prepared standards
Improperly made
standards
Recalibrate using freshly
made standards
Temperature
differences between
calibration and
sample
measurement
Ensure calibration and
sample measurements
are done at the same
temperature
No TISAB used Add TISAB buffer
Incorrect pH Ensure pH is between
5 -7
Variations in
measurement
procedure
between
calibration and
measurement
Ensure stirring rates,
temperature, and other
conditions are the same
in calibration and
measurement steps
Parameter Operating Range
Concentration range 0.02 to 19,000 mg/L
Temperature range 0 – 60ºC
pH range 5 -7 pH
Reference type Double Junction Ag/AgCl/rellable
Sensor materials Solid state crystal
Body and sleeve Polypropylene/PMMA
Overall length 150mm
Barrel diameter 12mm
Cable length 1m standard, longer to order.
Maximum 20m
Connector BNC standard, others on request
All specications and values are subject to change without notice. © 2012
Images are illustrative only.
GOOD CHEMISTRY
GOOD CHEMISTRY
Operators Manual
Short-Form
Intermediate Junction Series
Page view 0
1 2

Summary of Contents

Page 1

WarrantyIJ electrodes have a warranty of 12 months from date of purchase. Any electrode found to be faulty due to manufacture within this time will be

Page 2 - GOOD CHEMISTRY

Application The Ionode IJ-F Ion selective electrode responds reversibly to Fluoride ions. The limit of detection is 0.02 ppm F-. The response is theo

Comments to this Manuals

No comments